Vedere la pagina 70
2-39
ALARM OVERVIEW
ALARMSALARMS
ALARMSALARMS
ALARMS
ALARM OVERVIEW
This section covers alarms that may sound during testing
procedures. It includes a list of possible alarms, explanations
and the actions taken by the system.
If an alarm is sounded during operation of the unit, an Alarm
Screen will appear. For example, if a bubble is detected in
the fluid flow, the following screen will appear, an audio alarm
will sound and the test will be terminated.
CHANNEL 1
STOPPED-BUBBLE FOUND
Press to silence the audio portion of the alarm.
Press again to clear the visual portion of the alarm and
return to the IDLE or VIEW screen associated with the mode
the unit was working in when detecting the alarm state.
Most alarm states terminate the tests because further data
could be invalid. The alarm state should be cleared and the
test rerun. The data collected prior to the alarm is still recorded
in the datalogger for that test. If an alarm only pauses a test,
the system will autostart when the alarm state is cleared.
Vedere la pagina 70