Unlike all conventional ESU Analyzers with lesser degrees of accuracy, the ESU-2050
Series utilizes an external Current Transformer and external precision load resistors
(values to be determined by the manufacturer’s suggested test load for the generator you
are testing or servicing) for typical Electrosurgical Generator testing. Many of the world’s
leading Electrosurgical generator manufacturers utilize this exact same technique when
they test, service and calibrate their generators.
This methodology has several distinct advantages over conventional ESU Analyzers:
Improved accuracy and resolution capabilities
100% manufacturer recommended test load compatibility
Smaller and lighter weight instrumentation
The current transformer senses the RF current flowing through the external test load and
produces a proportional voltage as an input to the ESU-2050. This input is either a direct
1 : 1 input (for current transformers with a ratio of 1 : 1 volts : amps) or a 0.1 : 1 input (for
current transformers with a ratio of 0.1 : 1 volts : amps). Combining the standard and low
ranges of the ESU-2050 with the use of either a 0.1:1 or 1:1 ratio current transformer, the
user has full control over the ability to get high accuracy and high resolution readings from
all types of Electrosurgical Generators.